Electrode Placement and Audiologic Outcomes With the Advanced Bionics HiFocus L23 Electrode

The safety and scientific validity of this study is the responsibility of the study sponsor and investigators. Listing a study does not mean it has been evaluated by the U.S. Federal Government. Read our disclaimer for details.
 
ClinicalTrials.gov Identifier: NCT03858647
Recruitment Status : Active, not recruiting
First Posted : March 1, 2019
Last Update Posted : January 25, 2021
Sponsor:
Information provided by (Responsible Party):
Robert F. Labadie, Vanderbilt University Medical Center

Tracking Information
First Submitted Date February 12, 2019
First Posted Date March 1, 2019
Last Update Posted Date January 25, 2021
Actual Study Start Date February 21, 2019
Estimated Primary Completion Date December 2021   (Final data collection date for primary outcome measure)
Current Primary Outcome Measures
 (submitted: March 14, 2019)
  • Treatment Effectiveness-Consonant Nucleus Consonant (CNC) word test [ Time Frame: 2 years ]
    Audiologic speech performance as measured by CNC word recognition. CNC word tests consist of a 50 word list. Subjects repeat as much of each word as possible. Word scores will be calculated as percent correct at baseline through 24 months and reported as percentage points improved.
  • Treatment Effectiveness-AzBio sentence test [ Time Frame: 2 years ]
    Audiologic speech performance as measured by AzBio sentence recognition. AzBio consists of 20 sentence list spoken by different talkers (two male, two female). Subjects repeat as much of each sentence as possible. Each word correctly repeated counts towards the overall score, reported as an overall percentage correct at baseline through 24 months.
  • Treatment Effectiveness-Air-conduction thresholds [ Time Frame: 2 years ]
    Audiologic speech performance as measured by air-conduction thresholds. Threshold values will be obtained for 125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, and 8000 Hz using insert earphones. Post-operative hearing thresholds will be compared to pre-operative hearing thresholds to assess amount of residual hearing. Greater post-operative thresholds is associated with poorer hearing preservation.
Original Primary Outcome Measures
 (submitted: February 26, 2019)
  • Treatment Effectiveness-CNC word test [ Time Frame: 2 years ]
    Audiologic speech performance as measured by Consonant Nucleus Consonant (CNC) word recognition. CNC word tests consist of a 50 word list. Subjects repeat as much of each word as possible. Word scores will be calculated as percent correct at baseline through 24 months and reported as percentage points improved.
  • Treatment Effectiveness-AzBio sentence test [ Time Frame: 2 years ]
    Audiologic speech performance as measured by AzBio sentence recognition. AzBio consists of 20 sentence list spoken by different talkers (two male, two female). Subjects repeat as much of each sentence as possible. Each word correctly repeated counts towards the overall score, reported as an overall percentage correct at baseline through 24 months.
  • Treatment Effectiveness-Air-conduction thresholds [ Time Frame: 2 years ]
    Audiologic speech performance as measured by air-conduction thresholds. Threshold values will be obtained for 125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, and 8000 Hz using insert earphones. Post-operative hearing thresholds will be compared to pre-operative hearing thresholds to assess amount of residual hearing. Greater post-operative thresholds is associated with poorer hearing preservation.
Change History
Current Secondary Outcome Measures
 (submitted: March 14, 2019)
  • Treatment Effectiveness [ Time Frame: 2 years ]
    Determine scalar location after electrode insertion using intraoperative or postoperative imaging to then correlate imaging with audiologic outcomes as collected above.
  • Treatment Effectiveness - Audiologic outcomes [ Time Frame: 2 years ]
    Percentage of residual acoustic hearing for each subject as measured by the audiologic outcomes as collected above.
  • QOL Tinnitus Handicap Index (THI) [ Time Frame: 2 years ]
    Patient quality of life assessment: tinnitus compared post-implant to pre-implant. The sum of all responses to 25 questions is the THI score, with a range from 0-100. The higher the score, the more handicap the subject has due to tinnitus.
  • QOL Speech, Spatial and Qualities of Hearing scale (SSQ 12) [ Time Frame: 2 years ]
    Patient quality of life assessment: hearing and listening ability compared post implant to pre-implant. The SSQ 12 survey uses a visual analog rating scale from 0-10 over three listening domains: speech understanding in various listening conditions; spatial hearing associated with distance, movement, and direction; and the overall quality of speech including clarity and naturalness of sound. The higher the score, the better speech understanding, spatial hearing, and sound quality a subject has.
  • QOL Nijmegen [ Time Frame: 2 years ]
    Patient quality of life assessment: hearing ability scores compared post-implant to pre-implant. The Nijmegen is a 60 question survey covering three general domains: physical, psychological, and social functioning. Subjects rate each question on a scale of how often they can hear an event as noted. The subdomain mean score for each subject is derived by assigning values to survey items as follows: 1 = 0, 2 = 25, 3 = 50, 4 = 75, and 5 = 100 and then calculating the quantitative mean scores across the subdomains.
  • Evaluation of intraoperative ease of electrode insertion as assessed by surgeon. [ Time Frame: 2 years ]
    Surgeon survey
Original Secondary Outcome Measures
 (submitted: February 26, 2019)
  • Treatment Effectiveness [ Time Frame: 2 years ]
    Determine scalar location after electrode insertion using intraoperative or postoperative imaging to then correlate imaging with audiologic outcomes as collected above.
  • Treatment Effectiveness - Audiologic outcomes [ Time Frame: 2 years ]
    Percentage of residual acoustic hearing for each subject as measured by the audiologic outcomes as collected above.
  • QOL THI [ Time Frame: 2 years ]
    Patient quality of life assessment: tinnitus compared post-implant to pre-implant. The sum of all responses to 25 questions is the THI score, with a range from 0-100. The higher the score, the more handicap the subject has due to tinnitus.
  • QOL SSQ 12 [ Time Frame: 2 years ]
    Patient quality of life assessment: hearing and listening ability compared post implant to pre-implant. The SSQ 12 survey uses a visual analog rating scale from 0-10 over three listening domains: speech understanding in various listening conditions; spatial hearing associated with distance, movement, and direction; and the overall quality of speech including clarity and naturalness of sound. The higher the score, the better speech understanding, spatial hearing, and sound quality a subject has.
  • QOL Nijmegen [ Time Frame: 2 years ]
    Patient quality of life assessment: hearing ability scores compared post-implant to pre-implant. The Nijmegen is a 60 question survey covering three general domains: physical, psychological, and social functioning. Subjects rate each question on a scale of how often they can hear an event as noted. The subdomain mean score for each subject is derived by assigning values to survey items as follows: 1 = 0, 2 = 25, 3 = 50, 4 = 75, and 5 = 100 and then calculating the quantitative mean scores across the subdomains.
  • Evaluation of intraoperative ease of electrode insertion as assessed by surgeon. [ Time Frame: 2 years ]
    Surgeon survey
Current Other Pre-specified Outcome Measures Not Provided
Original Other Pre-specified Outcome Measures Not Provided
 
Descriptive Information
Brief Title Electrode Placement and Audiologic Outcomes With the Advanced Bionics HiFocus L23 Electrode
Official Title Multi-Center Assessment of Electrode Placement and Audiologic Outcomes in Cochlear Implantation With the Advanced Bionics HiFocus L23 Electrode
Brief Summary

The goal of this study is to measure patient performance after cochlear implantation with the Food and Drug Administration (FDA) approved Advanced Bionics HiFocus L23 device.

The specific aims of the project are as follows:

  1. To measure implant audiologic performance as defined by speech performance in standard cochlear implant speech test batteries.
  2. To determine scalar location after insertion through either intraoperative or postoperative imaging and correlate this with audiologic outcomes.
  3. To assess the rates of preservation of residual acoustic hearing and correlate those with scalar location as determined in specific aim 2 and audiologic performance as determined in specific aim 1.
  4. To correlate speech outcomes with quality of life measures, as defined by validated questionnaires including the Tinnitus Handicap Index (THI), Speech and Spatial Qualities (SSQ) Questionnaire, and Nijmegen Questionnaire.
  5. To evaluate intraoperative ease of insertion from the surgeon's perspective and correlate this with postoperative electrode location and audiologic performance.
Detailed Description

Cochlear implants (CI) are biomechanical sense prostheses approved by the FDA for the treatment of sensorineural hearing loss. CI's function by bypassing the absent or dysfunctional cochlear end organs to directly electrically stimulate the spiral ganglion and cochlear nerve. A variety of commercially available FDA-approved implants exist for the treatment of severe to profound deafness. Implants are typically designed in two broad categories: lateral wall electrodes and perimodiolar. Perimodiolar electrodes are precurved electrodes that are designed to hug the center of the cochlea (modiolus) after insertion. Lateral wall electrodes are slim and straight, hugging the lateral wall on insertion.

Recent studies have shown that electrode design and location within the cochlea impact patient performance. Specifically, electrodes located completely within the scala tympani are associated with better outcomes than those that cross over from the scala tympani to the scala vestibuli during insertion. The same study has shown that lateral wall electrodes are less likely than perimodiolar electrodes to have scalar excursion to the scala vestibuli. Location matters with these implants because electrode position completely within the scala tympani is associated with higher rates of preservation of residual hearing. Minimizing cochlear trauma allows for preservation of residual acoustic hearing particularly in low frequencies. While this low frequency hearing is of little benefit independently, when combined with the electrical stimulation from the cochlear implant patients have improved performance.

The HiFocus L23 electrode is manufactured by Advanced Bionics (Valencia, California), and is FDA approved for the treatment of sensorineural hearing loss in adult and pediatric patients. It is designed as a slim and straight lateral wall electrode that can be inserted with low force to minimize intracochlear trauma. While pre-market studies were performed by Advanced Bionics as a part of its approval for use by the FDA, no post-market studies have been performed to assess the audiologic outcomes, electrode location after insertion, and hearing preservation rates when using this electrode. These aims will be accomplished through several means. Audiologic speech performance will be assessed through a standardized clinical/audiometric protocol currently followed by the Vanderbilt University Medical Center cochlear implant team. Electrode location for all participants will be determined using novel software created by Vanderbilt engineering faculty and graduate students that uses computerized tomography (CT) images obtained as a part of practice postoperatively or intraoperatively using an FDA approved Xoran xCAT® device, a portable flat panel volume computerized tomography (fpVCT) machine (Vanderbilt) or regular CT scan at participating institutions as applicable. These images can be registered to existing pre-operative conventional CT scans and used to produce 3-dimensional models of electrode placement within each patient's cochlea. This software is investigational and used in other research studies for the same purpose.

The extent to which the electrode remains in the desired location (scala tympani) will then be correlated with audiologic outcomes as obtained above. Residual hearing rates and quality of life measures will be obtained as a part of audiologic testing.

If a fine resolution, high quality conventional head CT has been previously obtained, or is obtained as part of routine care this scan may be used in place of the fpVCT or in addition to the fpVCT scan. The option of having a conventional head CT performed in radiology for research purposes in place of the fpVCT head CT scan will also be available.

Surgeons will complete surveys after each case concludes to evaluate the ease of insertion and will record the insertion via standard recording techniques in the operating room that are available for any microscopic case. These videos for all surgeries will be blinded and reviewed at Vanderbilt for ease and speed of insertion.

Study Type Observational
Study Design Observational Model: Cohort
Time Perspective: Prospective
Target Follow-Up Duration Not Provided
Biospecimen Not Provided
Sampling Method Non-Probability Sample
Study Population All clinically confirmed sensorineural hearing loss patients who meet criteria for cochlear implantation using the Advanced Bionics HiFocus L23 electrode are eligible for enrollment.
Condition Sensorineural Hearing Loss
Intervention Not Provided
Study Groups/Cohorts Sensorineural hearing loss patients post cochlear implant
Patients who have documented sensorineural hearing loss and have received cochlear implantation (per standard of care).
Publications * Not Provided

*   Includes publications given by the data provider as well as publications identified by ClinicalTrials.gov Identifier (NCT Number) in Medline.
 
Recruitment Information
Recruitment Status Active, not recruiting
Estimated Enrollment
 (submitted: February 26, 2019)
100
Original Estimated Enrollment Same as current
Estimated Study Completion Date December 2022
Estimated Primary Completion Date December 2021   (Final data collection date for primary outcome measure)
Eligibility Criteria

Inclusion Criteria:

  • Male or female 18 years of age and above.
  • Sensorineural hearing loss meeting criteria for cochlear implants, assessed no greater than three months prior to cochlear implant surgery
  • Patient schedule to undergo CI surgery using Advanced Bionics SlimJ/L23 device.
  • Ability to undergo adequate audiology follow-up or planned follow up for those participants undergoing scanning in the OR.
  • Competent in decision making process and able to read and sign a written informed consent form.
  • English language proficiency.
  • If the patient has already had a fine resolution CT or a fpVCT of the head (more than likely as a participant of another research study) the scan will not be repeated, the images from the previous scan will be used for the purposes of this study.
  • If the patient has undergone a conventional CT scan of the head after cochlear implantation as part of their routine care, this scan may be used in place of the research scan or in addition to the research scan.

Exclusion Criteria:

  • Patients who weigh over 300 lbs (when Xoran xCAT headboard is being used). Otherwise, weight is not exclusionary.
  • Females who are pregnant. All females of child-bearing age in the study will undergo a pregnancy test.
Sex/Gender
Sexes Eligible for Study: All
Ages 18 Years and older   (Adult, Older Adult)
Accepts Healthy Volunteers No
Contacts Contact information is only displayed when the study is recruiting subjects
Listed Location Countries United States
Removed Location Countries  
 
Administrative Information
NCT Number NCT03858647
Other Study ID Numbers 181904
Has Data Monitoring Committee No
U.S. FDA-regulated Product
Studies a U.S. FDA-regulated Drug Product: No
Studies a U.S. FDA-regulated Device Product: No
IPD Sharing Statement
Plan to Share IPD: Undecided
Responsible Party Robert F. Labadie, Vanderbilt University Medical Center
Study Sponsor Vanderbilt University Medical Center
Collaborators Not Provided
Investigators
Principal Investigator: Alejandro Rivas, MD Vanderbilt University Medical Center