Working…
ClinicalTrials.gov
ClinicalTrials.gov Menu
Trial record 3 of 619 for:    oximeter

Performance of Pulse Oximeter Sensors in Neonates

The safety and scientific validity of this study is the responsibility of the study sponsor and investigators. Listing a study does not mean it has been evaluated by the U.S. Federal Government. Know the risks and potential benefits of clinical studies and talk to your health care provider before participating. Read our disclaimer for details.
 
ClinicalTrials.gov Identifier: NCT03870373
Recruitment Status : Recruiting
First Posted : March 12, 2019
Last Update Posted : September 12, 2019
Sponsor:
Collaborator:
University of California, San Francisco
Information provided by (Responsible Party):
Taiwan Aulisa Medical Devices Technologies, Inc.

Brief Summary:
Validate pulse oximeter sensors in neonates by comparing sensor readings to blood samples during cardiac surgery.

Condition or disease Intervention/treatment
Cardiac Surgery Device: Pulse Oximeter

Detailed Description:
Researchers at the University of California, San Francisco have developed methods that permit them to test multiple pulse oximeter sensors on neonates simultaneously during cardiac surgery. During this type of surgery, the arterial oxygen saturation changes over a wide range, providing an ideal arena for the assessment of neonatal pulse oximeter sensors at many levels of oxygen saturation. Several arterial blood gas samples are typically obtained during these procedures. Saturation of each arterial blood sample is determined by direct oximetry in a hemoximeter and this measurement is compared to the reading displayed on the pulse oximeter.

Layout table for study information
Study Type : Observational
Estimated Enrollment : 20 participants
Observational Model: Case-Only
Time Perspective: Prospective
Official Title: Validate Pulse Oximeter Sensors in Neonates by Comparing Sensor Readings to Blood Samples During Cardiac Surgery
Actual Study Start Date : February 1, 2019
Estimated Primary Completion Date : January 10, 2020
Estimated Study Completion Date : January 10, 2020

Resource links provided by the National Library of Medicine

MedlinePlus related topics: Heart Surgery

Group/Cohort Intervention/treatment
Test Subject
neonatal patients undergoing complex cardiac surgical procedures
Device: Pulse Oximeter
Aulisa's non-invasive pulse oximeter sensor is applied to a foot of the neonate; an adhesive patch secures the sensor to the neonate.




Primary Outcome Measures :
  1. Performance of Pulse Oximeter Sensor in Neonates [ Time Frame: Data are collected anywhere from the beginning of surgery until the completion of the procedure ]
    Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013)



Information from the National Library of Medicine

Choosing to participate in a study is an important personal decision. Talk with your doctor and family members or friends about deciding to join a study. To learn more about this study, you or your doctor may contact the study research staff using the contacts provided below. For general information, Learn About Clinical Studies.


Layout table for eligibility information
Ages Eligible for Study:   up to 1 Year   (Child)
Sexes Eligible for Study:   All
Accepts Healthy Volunteers:   No
Sampling Method:   Non-Probability Sample
Study Population
Neonatal patients undergoing complex cardiac surgical procedures.
Criteria

Inclusion Criteria:

  • Neonates scheduled to undergo cardiac surgery where routine blood draws would be taken
  • Consent signed by at least one legal guardian

Exclusion Criteria:

  • Presence of any skin irritation or breakdown on the foot
  • Foot impediments which would preclude proper placement of the sensor
  • Clinical condition precluding the use of adhesive materials

Information from the National Library of Medicine

To learn more about this study, you or your doctor may contact the study research staff using the contact information provided by the sponsor.

Please refer to this study by its ClinicalTrials.gov identifier (NCT number): NCT03870373


Contacts
Layout table for location contacts
Contact: Hung G Nguyen, MD (415) 476-7021 Hung.Nguyen@ucsf.edu

Locations
Layout table for location information
United States, California
UCSF Medical Center at Mission Bay Recruiting
San Francisco, California, United States, 94158
Contact: Hung G Nguyen, MD    415-476-7021    Hung.Nguyen@ucsf.edu   
Sponsors and Collaborators
Taiwan Aulisa Medical Devices Technologies, Inc.
University of California, San Francisco
Investigators
Layout table for investigator information
Principal Investigator: Hung G Nguyen, MD UCSF Medical Center at Mission Bay

Layout table for additonal information
Responsible Party: Taiwan Aulisa Medical Devices Technologies, Inc.
ClinicalTrials.gov Identifier: NCT03870373     History of Changes
Other Study ID Numbers: 18-25904
First Posted: March 12, 2019    Key Record Dates
Last Update Posted: September 12, 2019
Last Verified: September 2019
Individual Participant Data (IPD) Sharing Statement:
Plan to Share IPD: Undecided

Layout table for additional information
Studies a U.S. FDA-regulated Drug Product: No
Studies a U.S. FDA-regulated Device Product: Yes
Device Product Not Approved or Cleared by U.S. FDA: Yes
Pediatric Postmarket Surveillance of a Device Product: No
Product Manufactured in and Exported from the U.S.: No