Closed-Loop Control in Young Children 5-8 Years Old Using DiAs Platform

The safety and scientific validity of this study is the responsibility of the study sponsor and investigators. Listing a study does not mean it has been evaluated by the U.S. Federal Government. Read our disclaimer for details.
ClinicalTrials.gov Identifier: NCT02750267
Recruitment Status : Completed
First Posted : April 25, 2016
Results First Posted : June 23, 2017
Last Update Posted : July 24, 2017
Sponsor:
Collaborator:
DexCom, Inc.
Information provided by (Responsible Party):
Mark DeBoer, MD, University of Virginia

Brief Summary:
The overall aim of this proposed research is to determine the safety, feasibility and efficacy (AP vs at home use of SAP) of the Diabetes Assistant (DiAs) controller in day and night closed-loop control in young children 5-8 years old with type 1 diabetes over multiple 48 hours in an out-patient setting.

Condition or disease Intervention/treatment Phase
Diabetes Mellitus, Type 1 Device: Diabetes Assistant (DiAs) with Closed-Loop Not Applicable

Detailed Description:
Young children with Type 1 Diabetes (T1D) in the age range of 5-8 years old are a population with clear needs but unique challenges regarding the application of artificial pancreas (AP) technologies. Young children are likely to benefit from an AP system, with current deficits in glycemic control that include both significant hypoglycemia and sub-optimal HbA1c levels; however, they have undeveloped abilities to control and interact with the AP system, posing potential safety issues. During the hours that these children are away from their parents at school and elsewhere, they lack the sophistication to operate the currently-available tools in an AP system--and may induce harm if they are allowed to do so, causing parental resistance to AP use. Commercially-available insulin pumps have mechanisms to lock access to children to prevent inappropriate insulin-delivery. However, the AP is more complex than an insulin pump, both in requiring more detailed setting information (that a child could adversely alter) and in providing alerts for impending low- and high-blood glucose (BG) levels (that one wouldn't want to lock out to child use). These functions are all run via a platform on a smart phone—a device with which young children may already feel a high degree of familiarity and thus be more likely to attempt to explore and potentially change settings. It is likely that young children will benefit the most from a system that gives them access to some AP features but provides access to other features only for their parents. In this sense, young children require a device that is not user-centered as much as family-centered. A redesign of the system to provide appropriate access to AP tools—in which certain users can obtain access to certain functionalities—is direly needed before children in this age range can benefit from the improvements in blood glucose (BG) control that the AP has to offer.

Layout table for study information
Study Type : Interventional  (Clinical Trial)
Actual Enrollment : 12 participants
Allocation: Randomized
Intervention Model: Parallel Assignment
Masking: None (Open Label)
Primary Purpose: Treatment
Official Title: Closed-Loop Control in Young Children 5-8 Years Old Using DiAs Platform
Study Start Date : May 2016
Actual Primary Completion Date : May 2016
Actual Study Completion Date : May 2016

Resource links provided by the National Library of Medicine

MedlinePlus related topics: Diabetes Type 1

Arm Intervention/treatment
Experimental: Group A
In this randomized, cross-over study, the intervention involves blood glucose control using a Closed Loop system run by the Diabetes Assistant (DiAs) during a stay at a Research House/Hotel. All subjects will have blood glucose data compared between their usual diabetes care (at home, using home insulin pump) and this Closed-Loop care (at Research House/Hotel using the DiAs system). Subjects who are randomized to Group A will have home care evaluated before the Research House/Hotel admission. Subjects in this arm will participate in CGM training and data collection prior to the Research House/Hotel admission.
Device: Diabetes Assistant (DiAs) with Closed-Loop
All subjects will use DiAs Medical Platform, a study insulin pump and continuous glucose monitor (CGM) in closed-loop mode during a Research House/Hotel admission that will last up to 72 hours.

Experimental: Group B
Group B is identical to Group A with the exception that usual diabetes care (at home, using home insulin pump) will be evaluated after the Research House/Hotel admission. Subjects who are randomized to Group B will participate in CGM training and data collection after the Research House/Hotel admission. As with Group A, all subjects will use Diabetes Assistant (DiAs) with Closed-Loop during the admission.
Device: Diabetes Assistant (DiAs) with Closed-Loop
All subjects will use DiAs Medical Platform, a study insulin pump and continuous glucose monitor (CGM) in closed-loop mode during a Research House/Hotel admission that will last up to 72 hours.




Primary Outcome Measures :
  1. Percent of Sensor Glucose Readings Between 70-180 mg/dL [ Time Frame: 68 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.


Secondary Outcome Measures :
  1. Percent of Time Sensor Glucose Readings Are <70 mg/dL [ Time Frame: 68 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  2. Percent of Time Sensor Glucose Readings Are >150 mg/dL [ Time Frame: 72 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  3. Percent of Time Sensor Glucose Readings Are >180 mg/dL [ Time Frame: 68 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  4. Percent of Time Sensor Glucose Readings Are >250 mg/dL [ Time Frame: 68 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  5. Percent of Time Sensor Glucose Readings Are >400 mg/dL [ Time Frame: 72 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  6. Distribution of Sensor and Meter Glucose Values (Maximum, Minimum, Median, Interquartile Range, Mean, Standard Deviation) [ Time Frame: 72 hours ]
    All subjects have CGM and handheld glucose meter output analyzed and compared between time on closed-loop system and time on usual care period.

  7. Distribution of Sensor and Meter Glucose Values (Maximum) [ Time Frame: 72 hours ]
    All subjects have CGM and handheld glucose meter output analyzed and compared between time on closed-loop system and time on usual care period.

  8. Distribution of Sensor and Meter Glucose Values (Minimum) [ Time Frame: 72 hours ]
    All subjects have CGM and handheld glucose meter output analyzed and compared between time on closed-loop system and time on usual care period.

  9. Distribution of Sensor and Meter Glucose Values (Median/Interquartile Range) [ Time Frame: 72 hours ]
    All subjects have CGM and handheld glucose meter output analyzed and compared between time on closed-loop system and time on usual care period.

  10. Mean BG (as Measured by CGM) [ Time Frame: 68 hours ]
    All subjects have CGM and handheld glucose meter output analyzed and compared between time on closed-loop system and time on usual care period.

  11. Hypoglycemia Area Under the Curve <60 [ Time Frame: 72 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  12. Hypoglycemia Area Under the Curve <70 mg/dL [ Time Frame: 72 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  13. Hyperglycemia Area Under the Curve >180 [ Time Frame: 72 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  14. Hyperglycemia Area Under the Curve >250 mg/dL [ Time Frame: 72 hours ]
    All subjects have CGM output analyzed and compared between time on closed-loop system and time on usual care period.

  15. Incidence of Hypoglycemia Per Subject, Defined by Handheld Meter Glucose <70 mg/dL [ Time Frame: 68 hours ]
    All subjects have hypoglycemia monitored by meter analysis and compared between time on closed-loop system and time on usual care period.

  16. End of Night Blood Glucose [ Time Frame: 72 hours ]
    All subjects have blood glucose evaluated upon rising (approximately 7 am) and compared between time on closed-loop system and time on usual care period.



Information from the National Library of Medicine

Choosing to participate in a study is an important personal decision. Talk with your doctor and family members or friends about deciding to join a study. To learn more about this study, you or your doctor may contact the study research staff using the contacts provided below. For general information, Learn About Clinical Studies.


Layout table for eligibility information
Ages Eligible for Study:   5 Years to 9 Years   (Child)
Sexes Eligible for Study:   All
Accepts Healthy Volunteers:   No
Criteria

Inclusion Criteria:

  • Clinical diagnosis of type 1 diabetes,

    • The diagnosis of type 1 diabetes is based on the investigator's judgment
    • C peptide levels and antibody determinations are not required
  • Daily insulin therapy for ≥ 12 months
  • Insulin pump therapy for ≥ 3 months
  • Age ≥5 - ≤8 years old
  • Avoidance of acetaminophen-containing medications (i.e. Tylenol) while wearing the continuous glucose monitor.
  • Willingness to wear a continuous glucose sensor and physiological monitor for the duration of the study

Exclusion Criteria:

The presence of any of the following is an exclusion for the study:

  • Diabetic ketoacidosis in the past month
  • Hypoglycemic seizure or loss of consciousness in the past 3 months
  • History of seizure disorder (except for hypoglycemic seizure)
  • History of any heart disease including coronary artery disease, heart failure, or arrhythmias
  • Cystic fibrosis
  • Current use of oral glucocorticoids, beta-blockers or other medications, which in the judgment of the investigator would be a contraindication to participation in the study.
  • History of ongoing renal disease (other than microalbuminuria).
  • Subjects requiring intermediate or long-acting insulin (such as NPH, Detemir or Glargine).
  • Subjects requiring other anti-diabetic medications other than insulin (oral or injectable).
  • Presence of a febrile illness within 24 hours of admission or acetaminophen use while wearing the CGM. The subject may be rescheduled for Research House/Hotel Admission if these criteria are not met. The study subject will not participate in the trial if these conditions are met.
  • Medical or psychiatric condition that in the judgment of the investigator might interfere with the completion of the protocol such as:

    • Inpatient psychiatric treatment in the past 6 months
    • Uncontrolled adrenal insufficiency

Information from the National Library of Medicine

To learn more about this study, you or your doctor may contact the study research staff using the contact information provided by the sponsor.

Please refer to this study by its ClinicalTrials.gov identifier (NCT number): NCT02750267


Locations
Layout table for location information
United States, Virginia
University of Virginia Center for Diabetes Technology
Charlottesville, Virginia, United States, 22903
Sponsors and Collaborators
University of Virginia
DexCom, Inc.
Investigators
Layout table for investigator information
Principal Investigator: Mark D. DeBoer, MD University of Virginia, Pediatrics, Endocrinology/Diabetes

Additional Information:
Publications:
Layout table for additonal information
Responsible Party: Mark DeBoer, MD, Principal Investigator, University of Virginia
ClinicalTrials.gov Identifier: NCT02750267     History of Changes
Other Study ID Numbers: 18888
First Posted: April 25, 2016    Key Record Dates
Results First Posted: June 23, 2017
Last Update Posted: July 24, 2017
Last Verified: June 2017
Individual Participant Data (IPD) Sharing Statement:
Plan to Share IPD: Yes
Plan Description: Not yet determined

Keywords provided by Mark DeBoer, MD, University of Virginia:
Type 1 Diabetes Mellitus (T1DM)
Artificial Pancreas Project (APP)
Diabetes Assistant (DiAs)
Closed-Loop Control (CLC)
Continuous Glucose Monitor
Pediatric
Insulin Pump

Additional relevant MeSH terms:
Layout table for MeSH terms
Diabetes Mellitus
Diabetes Mellitus, Type 1
Glucose Metabolism Disorders
Metabolic Diseases
Endocrine System Diseases
Autoimmune Diseases
Immune System Diseases
Insulin
Hypoglycemic Agents
Physiological Effects of Drugs