Pulse Oximetry and Skin Pigmentation - Peds
Latest version (submitted March 15, 2022) on ClinicalTrials.gov
  • A study version is represented by a row in the table.
  • Select two study versions to compare. One each from columns A and B.
  • Choose either the "Merged" or "Side-by-Side" comparison format to specify how the two study versions are to be displayed. The Side-by-Side format only applies to the Protocol section of the study.
  • Click "Compare" to do the comparison and show the differences.
  • Select a version's Submitted Date link to see a rendering of the study for that version.
  • The yellow A/B choices in the table indicate the study versions currently compared below. A yellow table row indicates the study version currently being viewed.
  • Hover over the "Recruitment Status" to see how the study's recruitment status changed.
  • Study edits or deletions are displayed in red.
  • Study additions are displayed in green.
Study Record Versions
Version A B Submitted Date Changes
1 December 22, 2021 None (earliest Version on record)
2 March 15, 2022 Recruitment Status, Study Status, Contacts/Locations and Oversight
Comparison Format:

Scroll up to access the controls

Study NCT05185427
Submitted Date:  December 22, 2021 (v1)

Open or close this module Study Identification
Unique Protocol ID: STUDY00014644
Brief Title: Pulse Oximetry and Skin Pigmentation - Peds
Official Title: Characterizing Error in Pulse Oximetry by Comparing Skin Pigmentation With Pulse Oximeter and Arterial Line Reading Deviations in Intensive Care Unit Pediatric Patients
Secondary IDs:
Open or close this module Study Status
Record Verification: December 2021
Overall Status: Not yet recruiting
Study Start: February 2022
Primary Completion: February 2024 [Anticipated]
Study Completion: February 2024 [Anticipated]
First Submitted: December 22, 2021
First Submitted that
Met QC Criteria:
December 22, 2021
First Posted: January 11, 2022 [Actual]
Last Update Submitted that
Met QC Criteria:
December 22, 2021
Last Update Posted: January 11, 2022 [Actual]
Open or close this module Sponsor/Collaborators
Sponsor: University of Minnesota
Responsible Party: Sponsor
Collaborators:
Open or close this module Oversight
U.S. FDA-regulated Drug: No
U.S. FDA-regulated Device: No
Data Monitoring: No
Open or close this module Study Description
Brief Summary:

The investigators' study aims to study how melanin index (mx) affects the deviation between SpO2 and SaO2, which becomes generally greater as hypoxia increases. The studies reviewed grouped individuals by race or have assigned individuals into groups like "dark", "intermediate", or "light" to describe pigmentation. Both of these methods are neither standardized nor objective, looking for race identifiers when it is more useful to be considering skin pigmentation identifiers. Skin pigmentation is a spectrum and it should be treated as such when trying to characterize relationships involving measurable factors such as melanin index.

The investigators will similarly measure the deviation between SpO2 and SaO2 however novel in that the investigators will quantitatively measure skin pigmentation via a light reflectance measurement device by Photovault.

Detailed Description:

Pulse oximetry is an essential tool within healthcare due to its impressive ability to detect low oxygen levels even before hypoxia manifests in a patient. Despite being a powerful tool, pulse oximetry has limitations. It is well known that nail polish or jaundice can interfere with pulse oximetry readings. This knowledge influences clinical practice in that providers are aware of the inaccurately low saturation readings associated with nail polish and jaundice. Alarmingly, several studies have reported that pulse oximetry also fails to read as accurately for patients who have greater levels of skin pigmentation.

One theory reported by Bickler et al. (2005) and Fiener et al. (2007) suggests that pulse oximetry accuracy decreases as hypoxia increases, where this deviation is further amplified in the presence of greater levels of skin pigmentation. Bickler et al. (2005) reported statistically significant differences in pulse oximeter readings on three different pulse oximeters. They found a small bias existed at blood oxygen saturations above 80% and increased as desaturation increased. Specifically, this bias over-estimated blood oxygenation in Black patients where Bickler et al. reported a bias as much as 8% at very low oxygen saturation. A similar trend has been reported by others. Clinically, a staggering number of COVID-19 patients fall victim to "silent" or "apathetic" hypoxemia, exhibiting minimal symptoms of concern upon presentation, but rapidly experience multi-organ failure and in the most unfortunate circumstance, death, in less than 48 hours. If not concerned at the moment, emergency departments suggest patients self-monitor their oxygen saturation at home and return for further evaluation if they record an oxygenation reading below the COVID-19 protocol threshold. What would be a difference of a couple percentage points of observed oxygenation levels with a pulse oximeter in patients with more skin pigmentation determines admission, access to the required care, and risks a delay in the necessary oxygen supplementation of patients. One paper has even suggested that clinicians should accept 95% oxygen saturation for Black patients as opposed to a threshold of 92% SpO2 to ensure proper oxygenation. In January 2021, the World Health Organization included the use of pulse oximeters to assess and identify patients that would require hospitalization for COVID-19. However, in February 2021, the FDA cautioned against using pulse oximeter oxygen readings to diagnose or rule out COVID-19 patients, questioning its reliability in clinically significant circumstances shortly after the New England Journal of Medicine reported that Black patients suffered from occult hypoxemia nearly three times as much than White patients.

The investigators' study aims to study how melanin index (mx) affects the deviation between SpO2 and SaO2, which becomes generally greater as hypoxia increases. The studies reviewed grouped individuals by race or have assigned individuals into groups like "dark", "intermediate", or "light" to describe pigmentation. Both of these methods are neither standardized nor objective, looking for race identifiers when it is more useful to be considering skin pigmentation identifiers. Skin pigmentation is a spectrum and it should be treated as such when trying to characterize relationships involving measurable factors such as melanin index.

The investigators will similarly measure the deviation between SpO2 and SaO2 however novel in that the investigators will quantitatively measure skin pigmentation via a light reflectance measurement device by Photovault.

Open or close this module Conditions
Conditions: Pulse Oximetry
Keywords:
Open or close this module Study Design
Study Type: Observational
Observational Study Model: Case-Only
Time Perspective: Cross-Sectional
Biospecimen Retention:
Biospecimen Description:
Enrollment: 85 [Anticipated]
Number of Groups/Cohorts 1
Open or close this module Groups and Interventions
Groups/Cohorts Interventions
Participants with pulse oximetry readings
The research team will take a reflectance measurement on the same location that the most recent pulse oximetry measurement was taken. In addition, a reflectance measurement will be taken bilaterally for the nailbed, dorsal and ventral side of the digits, dorsal and ventral side of the hand, dorsal and ventral side of the forearm, lower leg, foot, and either side for the ear. The most recent pulse oximetry and arterial oxygen saturation readings will be accessed through the electronic medical records, those readings are generally conducted simultaneously and will provide comparison. The duration anticipated for an individual participant's participation is 1 day.
Open or close this module Outcome Measures
Primary Outcome Measures:
1. Difference between pulse oximetry saturation (SpO2) and arterial line saturation (SaO2)
[ Time Frame: 1 day ]

The reflective index will be measured in percent reflectance via the Photovault. The SpO2 is measured via pulse oximetry by the healthcare team as percent oxygen saturation. The SaO2 is measured via Arterial Blood Gas, which is also measured by the healthcare team. SaO2 is also reported in percent oxygen saturation. The difference between SpO2 and SaO2 as a factor of the reflective index will be assessed
Open or close this module Eligibility
Study Population: Pediatric Intensive Care Unit (ICU) patients, potentially including COVID-19 patients.
Sampling Method: Non-Probability Sample
Minimum Age:
Maximum Age: 17 Years
Sex: All
Gender Based:
Accepts Healthy Volunteers: No
Criteria:

Inclusion Criteria:

  • Currently being cared for in the University of Minnesota Masonic Children's Hospital Intensive Care Unit (ICU)
  • Already have an arterial line in place
  • ICU staff state that it would be a good time to approach possible participants

Exclusion Criteria:

  • If members of the research team are unable to access and measure with the Photovault the area that was measured via pulse oximetry by the care team
  • Patients who do not have SpO2 or SaO2 measurements
Open or close this module Contacts/Locations
Central Contact Person: Gwenyth Fischer, MD
Telephone: (612) 624-9574
Email: fisch662@umn.edu
Central Contact Backup: Anna Branstad
Telephone: (612) 626-2935
Email: brans104@umn.edu
Study Officials: Gwenyth Fischer, MD
Principal Investigator
University of Minnesota Medical School Department of Internal Medicine
Locations: United States, Minnesota
University of Minnesota
Minneapolis, Minnesota, United States, 55455
Open or close this module IPDSharing
Plan to Share IPD: No
Open or close this module References
Citations:
Links:
Available IPD/Information: