Pivotal Study of the Intravenous Blood Glucose (IVBG) System, In-Clinic Setting

This study has been completed.
Sponsor:
Information provided by:
DexCom, Inc.
ClinicalTrials.gov Identifier:
NCT00925080
First received: June 17, 2009
Last updated: February 2, 2010
Last verified: February 2010

June 17, 2009
February 2, 2010
June 2009
July 2009   (final data collection date for primary outcome measure)
IVBG System accuracy will primarily be assessed relative to ISO 15197 criteria (i.e., within ±15 mg/dL at YSI glucose levels < 75 mg/dL, and within ±20% at YSI glucose levels >75 mg/dL). [ Time Frame: 72-hours ] [ Designated as safety issue: No ]
Same as current
Complete list of historical versions of study NCT00925080 on ClinicalTrials.gov Archive Site
Not Provided
Not Provided
Not Provided
Not Provided
 
Pivotal Study of the Intravenous Blood Glucose (IVBG) System, In-Clinic Setting
Not Provided

The purpose of this study is to evaluate safety and accuracy of the IVBG System (the "System") when used to track blood glucose in insulin treated subjects with diabetes mellitus in an in-clinic setting for up to 72 hours (per subject). Reference blood glucose measurements will be collected across the entire reportable range of the System (e.g., 40 400 mg/dL) with adequate sampling at the upper and lower ends of this range. IVBG System accuracy will primarily be assessed relative to ISO 15197 criteria (i.e., within ±15 mg/dL at YSI glucose levels < 75 mg/dL, and within ±20% at YSI glucose levels >75 mg/dL).

Not Provided
Observational
Observational Model: Cohort
Time Perspective: Prospective
Not Provided
Not Provided
Non-Probability Sample

Individuals with insulin-requiring diabetes mellitus

  • Diabetes
  • Hyperglycemia
Device: IVBG
Venous blood glucose measurement every 7.5 minutes for 72-hours
A
Intervention: Device: IVBG
Not Provided

*   Includes publications given by the data provider as well as publications identified by ClinicalTrials.gov Identifier (NCT Number) in Medline.
 
Completed
50
August 2009
July 2009   (final data collection date for primary outcome measure)

Inclusion Criteria:

  1. Age 18 years or older;
  2. Willing to participate in one 72-hour in-clinic session during which up to 72 venous blood samples drawn for YSI measurement of glucose concentration (maximum sampling frequency of 4 per hour), a fingerstick blood glucose measurement will also be performed at the time of each YSI blood draw;
  3. If instructed, be willing not to perform any activities that would result in submersion of the Sensor/Potentiostat in water or willing to wear a waterproof covering when submerging the Sensor/Potentiostat;
  4. Have been diagnosed with insulin-requiring diabetes mellitus and are on multiple daily injections (MDI) or Continuous Subcutaneous Insulin Infusion (CSII) insulin therapy;
  5. Able to speak, read, and write English.

Exclusion Criteria:

  1. Have skin conditions or existing (or planned) medical instrumentation and/or dressings that preclude wearing the IVBG Sensor (e.g., extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis, or surgical dressings at the proposed wear site);
  2. Are pregnant (as demonstrated by a positive pregnancy test within 72-hours of device insertion);
  3. Have a contraindication to placement of a dedicated peripheral IV line;
  4. Have a known history of heparin-induced thrombocytopenia;
  5. Current participation in another investigational study protocol (if a subject has recently completed participation in another drug study, the subject must have completed that study at least 30 days prior to being enrolled in this study);
  6. Have any condition that, in the opinion of the Investigator, would interfere with their participation in the trial or pose an excessive risk to study staff handling venous blood samples (e.g., known history of hepatitis B or C).
Both
18 Years and older
No
Contact information is only displayed when the study is recruiting subjects
United States
 
NCT00925080
PTL-900193, Rev01
No
Andrew K. Balo / SVP, Clinical and Regulatory Affairs, and Quality Assurance, DexCom, Inc.
DexCom, Inc.
Not Provided
Not Provided
DexCom, Inc.
February 2010

ICMJE     Data element required by the International Committee of Medical Journal Editors and the World Health Organization ICTRP