Now Available for Public Comment: Notice of Proposed Rulemaking (NPRM) for FDAAA 801 and NIH Draft Reporting Policy for NIH-Funded Trials

Safety & Accuracy Study of the ActiSight Needle Guidance System in Patients Undergoing CT-Guided Percutaneous Chest Aspiration and Biopsy

This study has been completed.
Sponsor:
Information provided by:
ActiViews Ltd.
ClinicalTrials.gov Identifier:
NCT00819533
First received: January 8, 2009
Last updated: December 27, 2009
Last verified: December 2009

January 8, 2009
December 27, 2009
January 2008
October 2009   (final data collection date for primary outcome measure)
Frequency of target reached within a radius of 5 mm from the preplanned targeted point as measured on the final CT scan [ Time Frame: Measure evaluated during entire course of study ] [ Designated as safety issue: No ]
Frequency of target reached within a radius of 8 mm from the preplanned targeted point as measured on the final CT scan [ Time Frame: Measure evaluated during entire course of study ] [ Designated as safety issue: No ]
Complete list of historical versions of study NCT00819533 on ClinicalTrials.gov Archive Site
  • Number of punctures required to obtain sample [ Time Frame: Measure evaluated during entire course of study ] [ Designated as safety issue: No ]
  • Number of intermediate CT scans required to obtain sample [ Time Frame: During entire course of study ] [ Designated as safety issue: No ]
  • An adequate sample obtained during the procedure [ Time Frame: during entire course of study ] [ Designated as safety issue: No ]
  • Number of punctures required to obtain sample [ Time Frame: Measure evaluated during entire course of study ] [ Designated as safety issue: No ]
  • Number of intermediate CT scans required to obtain sample [ Time Frame: During entire course of study ] [ Designated as safety issue: No ]
Not Provided
Not Provided
 
Safety & Accuracy Study of the ActiSight Needle Guidance System in Patients Undergoing CT-Guided Percutaneous Chest Aspiration and Biopsy
A Prospective, Open-Label, Single-Arm, Single-Center Study Evaluating the Preliminary Safety and Accuracy of the ActiSight Needle Guidance System in Patients Undergoing CT-Guided Percutaneous Chest Aspiration and Biopsy

This study is designed to evaluate preliminary safety and accuracy of the ActiSight™ Needle Guidance System in patients undergoing CT-guided lung aspiration or biopsy. This image guidance system uses a combination of CT scans and camera images to help guide a biopsy needle to a lung lesion. The system displays a simulated image of a biopsy needle and its estimated path for the biopsy.

Why is this study conducted to demonstrate the safety and effectiveness of the ActiSight Needle Guidance System in assisting trained physicians perform CT guided percutaneous needle biopsy (PNB) and fine needle aspiration (FNA) procedures in the chest.

Subjects, who are scheduled for medically indicated chest PNB or FNA procedures and meet all the inclusion and exclusion criteria, will be asked to participate in the study. The aim is to enroll 40 subjects that will complete the whole study.

Recruited subjects will go through screening tests and if eligible, will undergo, within 18 days of screening, CT Guided chest PNB or FNA utilizing the ActiSight Needle Guidance system. Subjects will be observed and tested for at least an hour after the procedure to detect any serious complications. A follow-up will be conducted within 7-10 days after the procedure day for safety evaluations. Safety and accuracy data from all stages of the study will be collected and analyzed.

The effectiveness of the guidance system will be measured primarily by the frequency of reaching the intended targets, as determined by a final CT scan. Other measures will include the number of needle punctures and the number of intermediate CT scans needed to complete the procedure. The safety of the system will be measured by the frequency and severity of adverse events.

Procedure description:

First, a single-use, disposable pad (ActiSticker) is attached to the patient's skin over the point of needle insertion as selected by the interventionist. The pad incorporates reference objects that are easily detectable both on CT images and visually. A volume CT scan, covering at least the pad, the target lesion and adjacent structures is then performed. Next, the CT images from the volume scan are transferred to the ActiSight computer where the operator marks the target of the procedure, as well as all the reference points on the ActiSticker. Next, an optical sensor, also single-use and disposable, is attached to the top of the needle. By optically detecting all the reference points and relating them to their known positions on the CT images, the system can calculate the position of the top and the tip of the needle with relation to the target lesion. During the procedure, the system shows the user, on a graphic computer display, whether the needle is aimed directly at the target and how far the tip of the needle is from the intended target to help the user aim the insertion and guidance of the needle to the target.

Interventional
Not Provided
Allocation: Non-Randomized
Endpoint Classification: Safety/Efficacy Study
Intervention Model: Single Group Assignment
Masking: Open Label
Primary Purpose: Diagnostic
Biopsy
Device: sensor (ActiSight™ Needle Guidance System)
Patients will have their lung sample obtained under CT and ActiSight needle guidance system
Experimental: sensor
Patients will have their lung sample obtained under CT and ActiSight needle guidance system
Intervention: Device: sensor (ActiSight™ Needle Guidance System)
Not Provided

*   Includes publications given by the data provider as well as publications identified by ClinicalTrials.gov Identifier (NCT Number) in Medline.
 
Completed
40
October 2009
October 2009   (final data collection date for primary outcome measure)

Inclusion Criteria:

  • Male or female subjects, 18 years of age or older at the time of enrollment
  • Subjects meeting all medical conditions for percutaneous chest aspiration/biopsy with safe path to lesion between ribs, however, not in the breast area in women or men with gynecomastia
  • Single target pulmonary lesion greater than 10 mm in size (diameter) and depth of lesion greater than 40 mm from the skin
  • Written informed consent to participate in the study
  • Ability to comply with the requirements of the study procedures

Exclusion Criteria:

  • Fibrous tissue or cyst in access path (not including target lesion)
  • Hemophilia, thrombocytopenia, coagulopathy, or other elevated risk for bleeding or abnormal clotting
  • Use of ticlodipine or similar antithrombotic medication
  • For subjects taking warfarin or other anticoagulant medication, INR >1.3
  • Pulmonary hypertension, heart failure, renal failure, or blood dyscrasia
  • Patient is unable to comply with requirements of the procedure, i.e. holding breath
  • Participation in any prior investigational trial within 30 days of enrollment
  • Any form of substance abuse (including drug or alcohol abuse), psychiatric disorder, or any chronic condition that could, in the opinion of the investigator, of interfering with the conduct of the study.
  • Subjects who are uncooperative or cannot follow instructions
Both
18 Years and older
No
Contact information is only displayed when the study is recruiting subjects
Canada
 
NCT00819533
ASNG-LFNA-101-CAN
No
Uri Shreter, PhD / Study Director, ActiViews
ActiViews Ltd.
Not Provided
Principal Investigator: Narinder Paul, MD University Health Network, Toronto General Hospital
Study Director: Uri Shreter, Ph.D. ActiViews Ltd.
ActiViews Ltd.
December 2009

ICMJE     Data element required by the International Committee of Medical Journal Editors and the World Health Organization ICTRP