Performance Evaluation of Blood Glucose Monitoring Systems (Polaris)
- Full Text View
- Tabular View
- No Study Results Posted
- Disclaimer
- How to Read a Study Record
Purpose
The goal of this single-center, comparative, open label, in-vitro diagnostic device performance evaluation was to investigate the performance of two sanofi-aventis BGMSs (BGStar® and iBGStar®) with regard to system accuracy (Polaris), precision and Hct influence (Helios).
| Condition | Intervention | Phase |
|---|---|---|
|
Diabetes Mellitus |
Other: Sugar infusion Drug: insulin infusion |
Phase 4 |
| Study Type: | Interventional |
| Study Design: | Allocation: Non-Randomized Endpoint Classification: Efficacy Study Intervention Model: Single Group Assignment Masking: Open Label Primary Purpose: Diagnostic |
| Official Title: | Performance Evaluation of BGStar and iBGStar in Terms of Accuracy (Polaris), Intra-assay Precision, and Hematocrit Interference (Helios Substudy) Following ISO15197 and TNO Guidelines |
- Accuracy of the blood glucose meter in comparison to a standard reference [ Time Frame: within 10 min up to 300 min after start of experiment ] [ Designated as safety issue: No ]After start of the experiment, blood glucose will be measured by means of the test devices and blood will be drawn for assessment of venous blood glucose concentration with a standard laboratory reference method
- Intra-Assay precision [ Time Frame: within 10 min up to 300 min after start of experiment ] [ Designated as safety issue: No ]Blood samples from 5 individuals with different levels of blood glucose were each measured with 10 meters of both investigational devices, the BGStar® and iBGStar®, ten times resulting in a total number of 500 measurements.
- Hematocrit Interference (Helios) [ Time Frame: within 10 min up to 300 min after start of the experiment ] [ Designated as safety issue: No ]One patient provided blood and artificial samples are generated from venous heparinized blood. These blood samples were adjusted to five different BG levels (2.8 mmol/L, 5.6 mmol/L, 8.3 mmol/L, 13.9 mmol/L, 19.4 mmol/L) at five different Hct levels (35%, 40%, 45%, 50%, 55%) to fulfill the range required by the TNO 2001 and tested by all devices.
| Enrollment: | 106 |
| Study Start Date: | January 2011 |
| Study Completion Date: | June 2011 |
| Primary Completion Date: | April 2011 (Final data collection date for primary outcome measure) |
| Arms | Assigned Interventions |
|---|---|
|
Sugar infusion
Diagnostic assessment of blood glucose by means of different devices
|
Other: Sugar infusion
Infusion of glucose to achieve high blood glucose levels
Other Name: 5 % glucose solution
|
|
Experimental: insulin infusion
infusion of insulin to achieve low glucose levels
|
Drug: insulin infusion
i.v. infusion of insulin
Other Name: insuman rapid
|
Detailed Description:
The main study (Polaris) investigated blood glucose measuring accuracy and intra-assay precision while the substudy (Helios) investigated potential haematocrit interference of sanofi-aventis BGMSs BGStar® and iBGStar® and partly of competitor BGMSs according to ISO 15197 [1] and TNO 2001 Quality Guideline [2], respectively.
The primary objective of the system accuracy evaluation is to identify the number of blood glucose readings of both sanofi-aventis devices within the ISO 15197 area of acceptable accuracy of ± 20% [DEV%] for blood samples with glucose concentration > 75 mg/dL (4.2 mmol/L) and within the ISO 15197 area of acceptable accuracy of ± 15 mg/dL (± 0.83 mmol/L) for blood samples with glucose concentration < 75 mg/dL (4.2 mmol/L) as compared to the StatStrip® Connectivity reference method. The primary objective of the haematocrit interference evaluation is to demonstrate that the BGStar® and iBGStar® devices meet the Hct interference requirements according to TNO 2001 Quality Guideline when used to measure BG of venous whole blood as compared to a venous plasma reference method (YSI 2300 STAT Plus™) at Hct levels between 35 - 55% in terms of:(1) a mean deviation not exceeding ± 10% [DEV%] for hyperglycaemic glucose concentrations (150 - 180, 250 - 280, 320 - 350 mg/dL) and (2)a mean deviation not exceeding ± 18 mg/dL (1.0 mmol/L) [DEVtotal] for hypo- and normoglycaemic glucose values (50 - 60, 100 - 120 mg/dL) compared to the reference method at Hct levels between 35 - 55%.
Eligibility| Ages Eligible for Study: | 18 Years to 75 Years |
| Genders Eligible for Study: | Both |
| Accepts Healthy Volunteers: | Yes |
Inclusion Criteria:
- healthy subjects or patients with type 1 or type 2 diabetes
- screening glucose levels fall into required glucose ranges
- screening hematocrit values between 36 % to 55 %
Exclusion Criteria:
- history of hypotension during blood draws
- intake of drugs known to interfere with blood glucose readings
- biochemical safety parameters outside of reference ranges
- Hb < 11 g/dL
- lack of compliance
- history of frequent hypoglycemia
Contacts and Locations| Germany | |
| IKFE - Institute for Clinical Research and Development | |
| Mainz, Germany, 55116 | |
| Principal Investigator: | Andreas Pfützner, MD, PhD | IKFE Institute for Clinical Research and Development |
More Information
No publications provided
| Responsible Party: | IKFE Institute for Clinical Research and Development |
| ClinicalTrials.gov Identifier: | NCT01504620 History of Changes |
| Other Study ID Numbers: | BGSTA_C_05445, SAN-HCT-002 Helios, SAN-BGM-002 Precision |
| Study First Received: | December 27, 2011 |
| Last Updated: | January 3, 2012 |
| Health Authority: | Germany: Federal Institute for Drugs and Medical Devices |
Keywords provided by IKFE Institute for Clinical Research and Development:
|
diabetes mellitus blood glucose self-assessment Accuracy Intra-Assay precision Hematocrit interference |
Additional relevant MeSH terms:
|
Diabetes Mellitus Glucose Metabolism Disorders Metabolic Diseases Endocrine System Diseases |
Insulin Hypoglycemic Agents Physiological Effects of Drugs Pharmacologic Actions |
ClinicalTrials.gov processed this record on May 22, 2013