Performance of an Investigational Blood Glucose Monitoring System in a Clinical Setting
This study has been completed.
Sponsor:
Bayer HealthCare, Diabetes Care
Information provided by (Responsible Party):
Bayer HealthCare, Diabetes Care
ClinicalTrials.gov Identifier:
NCT01268488
First received: December 29, 2010
Last updated: September 14, 2012
Last verified: September 2012
- Full Text View
- Tabular View
- Study Results
- Disclaimer
- How to Read a Study Record
Purpose
The purpose of this study is to evaluate the performance of an investigational blood glucose meter with the Contour sensor strip. It was decided that this investigational monitoring system would not be part of a regulatory submission and it will not proceed to marketed product.
| Condition | Intervention |
|---|---|
|
Diabetes |
Device: Ninja 2 Investigational Blood Glucose Meter |
| Study Type: | Interventional |
| Study Design: | Endpoint Classification: Efficacy Study Intervention Model: Single Group Assignment Masking: Open Label Primary Purpose: Diagnostic |
| Official Title: | Clinical Trial Protocol for NINJA 2 System With CONTOUR Sensor |
Resource links provided by NLM:
Further study details as provided by Bayer HealthCare, Diabetes Care:
Primary Outcome Measures:
- Numbers of Fingerstick Blood Glucose (BG) Results Within +/- 5 to 15mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method [ Time Frame: 2 hours ] [ Designated as safety issue: No ]Subjects with diabetes tested subject fingerstick blood using an investigational blood glucose meter (BGM). BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5 to 20% (for reference BG results >=75mg/dL) of the reference method results.
Secondary Outcome Measures:
- Numbers of Palm Blood Glucose (BG) Results Within +/- 5 to 15 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method [ Time Frame: 2 hours ] [ Designated as safety issue: No ]Subjects with diabetes used an investigational blood glucose meter (BGM) with subject capillary blood obtained from the palm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results.
- Number of Subjects Rated as<=3 Performing Basic Meter Tasks (Labeling Comprehension) [ Time Frame: 2 hours ] [ Designated as safety issue: No ]Subjects reviewed the instructions for use (User Guide and Quick Reference Guide) to learn to use the system. Study staff then observed and rated the subjects(1 to 4) on their success at performing the tasks. Scale: 1.Success in performing tasks correctly without assistance. 2.Successful with additional review of User Guide. 3.Successful with additional review and study staff assist similar to review of a specific function during a Customer Service call. 4.Subject did not perform task correctly and study staff intervention was required.
- Numbers of Forearm Blood Glucose (BG) Results Within +/- 5 to 15 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method [ Time Frame: 2 hours ] [ Designated as safety issue: No ]Subjects with diabetes used an investigational blood glucose meter (BGM) with subject capillary blood obtained from the forearm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results.
| Enrollment: | 98 |
| Study Start Date: | December 2010 |
| Study Completion Date: | December 2010 |
| Primary Completion Date: | December 2010 (Final data collection date for primary outcome measure) |
| Arms | Assigned Interventions |
|---|---|
|
Experimental: Intended Users of the System
Untrained subjects with diabetes obtained capillary fingerstick, palm, and forearm blood and performed Blood Glucose (BG) tests using a Ninja 2 investigational blood glucose meter and the Contour® sensor. This BG monitoring system will not proceed to marketed product.
|
Device: Ninja 2 Investigational Blood Glucose Meter
Untrained subjects with diabetes performed Blood Glucose(BG) tests from the subject's capillary blood obtained from fingerstick, palm, and forearm using the Ninja 2 investigational meter. All BG results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff.
|
Eligibility| Ages Eligible for Study: | 18 Years and older |
| Genders Eligible for Study: | Both |
| Accepts Healthy Volunteers: | No |
Criteria
Inclusion Criteria:
- Males and females, 18 years of age and above
- Type 1 or type 2 diabetes
- Currently self-testing blood glucose at home at least twice daily
- Able to speak, read, and understand English
- Willing to complete all study procedures
Exclusion Criteria:
- Currently Pregnant
- Hemophilia or any other bleeding disorder
- Employee of competitive medical device company
- Cognitive disorder or other condition which, in the opinion of the investigator, would put the person at risk or seriously compromise the integrity of the study
Contacts and Locations
Please refer to this study by its ClinicalTrials.gov identifier: NCT01268488
Locations
| United States, New Jersey | |
| Consumer Product Testing Company | |
| Fairfield, New Jersey, United States, 07004 | |
Sponsors and Collaborators
Bayer HealthCare, Diabetes Care
Investigators
| Principal Investigator: | Michael Caswell, PhD | Consumer Product Testing Company |
More Information
No publications provided
| Responsible Party: | Bayer HealthCare, Diabetes Care |
| ClinicalTrials.gov Identifier: | NCT01268488 History of Changes |
| Other Study ID Numbers: | CTD-2010-008-02 |
| Study First Received: | December 29, 2010 |
| Results First Received: | September 14, 2012 |
| Last Updated: | September 14, 2012 |
| Health Authority: | United States: Institutional Review Board |
ClinicalTrials.gov processed this record on June 17, 2013