Performance of an Investigational Blood Glucose Monitoring System in a Clinical Setting

This study has been completed.
Sponsor:
Information provided by (Responsible Party):
Bayer HealthCare, Diabetes Care
ClinicalTrials.gov Identifier:
NCT01268488
First received: December 29, 2010
Last updated: September 14, 2012
Last verified: September 2012
  Purpose

The purpose of this study is to evaluate the performance of an investigational blood glucose meter with the Contour sensor strip. It was decided that this investigational monitoring system would not be part of a regulatory submission and it will not proceed to marketed product.


Condition Intervention
Diabetes
Device: Ninja 2 Investigational Blood Glucose Meter

Study Type: Interventional
Study Design: Endpoint Classification: Efficacy Study
Intervention Model: Single Group Assignment
Masking: Open Label
Primary Purpose: Diagnostic
Official Title: Clinical Trial Protocol for NINJA 2 System With CONTOUR Sensor

Resource links provided by NLM:


Further study details as provided by Bayer HealthCare, Diabetes Care:

Primary Outcome Measures:
  • Numbers of Fingerstick Blood Glucose (BG) Results Within +/- 5 to 15mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method [ Time Frame: 2 hours ] [ Designated as safety issue: No ]
    Subjects with diabetes tested subject fingerstick blood using an investigational blood glucose meter (BGM). BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5 to 20% (for reference BG results >=75mg/dL) of the reference method results.


Secondary Outcome Measures:
  • Numbers of Palm Blood Glucose (BG) Results Within +/- 5 to 15 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method [ Time Frame: 2 hours ] [ Designated as safety issue: No ]
    Subjects with diabetes used an investigational blood glucose meter (BGM) with subject capillary blood obtained from the palm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results.

  • Number of Subjects Rated as<=3 Performing Basic Meter Tasks (Labeling Comprehension) [ Time Frame: 2 hours ] [ Designated as safety issue: No ]
    Subjects reviewed the instructions for use (User Guide and Quick Reference Guide) to learn to use the system. Study staff then observed and rated the subjects(1 to 4) on their success at performing the tasks. Scale: 1.Success in performing tasks correctly without assistance. 2.Successful with additional review of User Guide. 3.Successful with additional review and study staff assist similar to review of a specific function during a Customer Service call. 4.Subject did not perform task correctly and study staff intervention was required.

  • Numbers of Forearm Blood Glucose (BG) Results Within +/- 5 to 15 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method [ Time Frame: 2 hours ] [ Designated as safety issue: No ]
    Subjects with diabetes used an investigational blood glucose meter (BGM) with subject capillary blood obtained from the forearm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results.


Enrollment: 98
Study Start Date: December 2010
Study Completion Date: December 2010
Primary Completion Date: December 2010 (Final data collection date for primary outcome measure)
Arms Assigned Interventions
Experimental: Intended Users of the System
Untrained subjects with diabetes obtained capillary fingerstick, palm, and forearm blood and performed Blood Glucose (BG) tests using a Ninja 2 investigational blood glucose meter and the Contour® sensor. This BG monitoring system will not proceed to marketed product.
Device: Ninja 2 Investigational Blood Glucose Meter
Untrained subjects with diabetes performed Blood Glucose(BG) tests from the subject's capillary blood obtained from fingerstick, palm, and forearm using the Ninja 2 investigational meter. All BG results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff.

  Eligibility

Ages Eligible for Study:   18 Years and older
Genders Eligible for Study:   Both
Accepts Healthy Volunteers:   No
Criteria

Inclusion Criteria:

  • Males and females, 18 years of age and above
  • Type 1 or type 2 diabetes
  • Currently self-testing blood glucose at home at least twice daily
  • Able to speak, read, and understand English
  • Willing to complete all study procedures

Exclusion Criteria:

  • Currently Pregnant
  • Hemophilia or any other bleeding disorder
  • Employee of competitive medical device company
  • Cognitive disorder or other condition which, in the opinion of the investigator, would put the person at risk or seriously compromise the integrity of the study
  Contacts and Locations
Choosing to participate in a study is an important personal decision. Talk with your doctor and family members or friends about deciding to join a study. To learn more about this study, you or your doctor may contact the study research staff using the Contacts provided below. For general information, see Learn About Clinical Studies.

Please refer to this study by its ClinicalTrials.gov identifier: NCT01268488

Locations
United States, New Jersey
Consumer Product Testing Company
Fairfield, New Jersey, United States, 07004
Sponsors and Collaborators
Bayer HealthCare, Diabetes Care
Investigators
Principal Investigator: Michael Caswell, PhD Consumer Product Testing Company
  More Information

No publications provided

Responsible Party: Bayer HealthCare, Diabetes Care
ClinicalTrials.gov Identifier: NCT01268488     History of Changes
Other Study ID Numbers: CTD-2010-008-02
Study First Received: December 29, 2010
Results First Received: September 14, 2012
Last Updated: September 14, 2012
Health Authority: United States: Institutional Review Board

ClinicalTrials.gov processed this record on August 28, 2014